2.D7 UI connectors ZED700-FS, ZED701-FS, ZED702-FS
D7 connectors are used, depending on connector type, to record voltages and currents in
three different variants and each of these at two rates; they do so by means of an integrated
24-bit delta-sigma A/D converter. At five measuring operations per second (mops)
200,000 digits can be acquired or at 500 mops, 20,000 digits. The necessary sensor supply
(U+) and (Gnd) can be set via the device itself at 6, 9, or 12 V. The 4-wire configuration
avoids measuring errors resulting from a voltage drop on Gnd.
Measuring quantities and ranges - factory default settings
Designation
M/s Quantity
Measuring range Units Resolution
ZED700-FS
1. * U2.00000
5
B-01
D U25
-2.2...+2.2
V
0.01 mV
2. U2.0000
5OO
B-02
D U24
-2.2...+2.2
V
0.1 mV
3. U2.000
1000
B-03
D U23
-2.2...+2.2
V
1 mV
ZED700-FS2
1. * U250.000
5
B-01
D U254
-250...+250
mV
1 uV
2. U64.000
5
B-02
D U643
-64...+64
mV
1 uV
ZED701-FS
1. * I20.0000
5
B-01
D I2O4
-20.0...+20.0
mA
0.1 uA
2. I20.000
5OO
B-02
D I2O3
-20.0...+20.0
mA
1 uA
3. I20.00
1000
B-03
D I202
-20.0...+20.0
mA
10 uA
ZED702-FS
1. * U20.0000
5
B-01
D U2O4
-20.0...+20.0
V
0.1 mV
2. U20.000
5OO
B-02
D U2O3
-20.0...+20.0
V
10 mV
3. U20.00
1000
B-03
D U2O2
-20.0...+20.0
V
10 mV
The preset measuring quantities and ranges are indicated by an asterisk (*).