5.2.Sensor adjustment and scaling
These two quantities offer two different conversion rates and thus also different measuring
ranges and resolutions. It is important to strictly adhere to each of the maximum final valu-
es including the decimal point.
For the purpose of zero adjustment the 'ZERO' key is provided.
For configuring the sensor's individual gain and scaling there are two possible procedures.
1. If the sensor was supplied with a specific 'characteristic value ' in mV / V, this can
easily be entered in the sensor menu; scaling will then be performed with the final
value and decimal point.
2. Or, alternatively, the indicator can be deleted and the sensor can be subjected to a
force equivalent to the final value. Scaling will then be performed, similarly,
using the final value and the decimal point. Pressing the 'ADJ' key will adjust the
gain accordingly.
For taring purposes the standard function for 'zero-setting' the measured value is available.
5.3.Pin assignment
Technical data
Power supply
from 6 V up, from the ALMEMO
®
device itself
Bridge voltage
5 V stabilized (terminals 5V and Gnd)
A/D converter
Delta-sigma, ratiometric (Bridge voltage = reference)
Common-mode range
0.25 to 4.75 V
Input range
-29.300 to +29.300 mV (terminals In+ and In-)
Display range
0 to ±200,000 (range DMS1 at 10 mops)
0 to ±50,000 (range DMS2 at 1000 mops)
Accuracy (10mops)
0.02% ±2 digits
Temperature drift
maximum 30 ppm/K
Nominal temperature
+22 °C ±2 K
Self-calibration
Each time the device is switched on or a sensor is connected
Refresh rate
DMS1 100 ms, DMS2 1 ms
Current consumption
approx. 15.5 mA
Connector colors
Housing ruby red, levers black
- 11 -