5.D7 bridge connector ZKD700-FS
D7 bridge connector ZKD700-FS is used to record the output voltage of a 4-wire full
bridge with a 5V supply (5V, Gnd); it does so by means of an integrated high-speed 24-
bit A/D converter. It is thus possible to record dynamic changes in force by means of
wire strain gauges at two different conversion rates, namely 10 or 1000 measuring opera-
tions per second. Connection is via four screw terminals.
Measuring quantities and ranges - factory default settings
Designation
Quantity
Measuring range Final value Units
Resolution
1. Kraft 1000mops
B-02
DMS2
0..+50000.
50000
xx
1 xx
Adjustment is performed by means of internal high-precision resistors each time the de-
vice is switched on, a sensor is connected, or there is a change in the quantity being mea-
sured. Depending on the specifications of the wire strain gauges the measuring channel
must be programmed with the required units. Scaling is performed via the sensor menu
either on the V7 measuring instrument or on the PC.
5.1.Configuration via the sensor menu
The following two measuring ranges can be configured. (* factory default settings)
Designation
Quantity
Measuring range Final value Units
Resolution
1. Kraft 10mops
B-01
DMS1
0..+200000.
200000
xx
1 xx
2. * Kraft 1000mops
B-02
DMS2
0..+50000.
50000
xx
1 xx