22. D6 V-lambda-radiation sensor FLAD03VL1
The D6 V-lamda radiation sensor FLAD03VL1 serves to measure the spectral range
of the visible light. The wavelength range extends from the end of the UV spectrum at
400 nm to the beginning of the IR range at 720 nm with a maximum at 555 nm. The
spectral sensitivity of the receiver is extremely well adapted to the sensitivity of the
human eye and complies with the device class B as per DIN 5032. The determined il-
luminance in “LUX” can directly be converted into the irradiance “W/ m²”. The AL-
MEMO
®
D6 sensor features 4 sensor channels: one for the kilolux range and three
other channels with various resolutions for the lux range.
22.1 Measuring ranges upon delivery
Designation
Range
Exp
Measuring range Dim Resolution
1. Ev kLux
B-01
DIGI
-2
0 to 200,00 kL
0,01 kLux
2. Ev Lux 0
B-02
DIGI
0
0 to 65000 Lx
1 Lux
3. Ev Lux 1
B-03
DIGI
-1
0 to 6500,0 Lx
0,1 Lux
4. Ev Lux 2
B-04
DIGI
-2
0 to 650,00 Lx
0,01 Lux
22.2 Configuration on the PC via the sensor menu
The measuring ranges of the measuring channels can be configured according to a list
of ranges (*factory default settings):
22.2.1 Configurable Measuring ranges
Designation
Range
Exp
Measuring range Dim Resolution
1. *Ev kLux
B-01
DIGI
-2
0 to 200,00 kL
0,01 kLux
2. *Ev Lux 0
B-02
DIGI
0
0 to 65000 Lx
1 Lux
3. *Ev Lux 1
B-03
DIGI
-1
0 to 6500,0 Lx
0,1 Lux
4. *Ev Lux 2
B-04
DIGI
-2
0 to 650,00 Lx
0,01 Lux
- 46 -