20.2.2 Measuring functions
To fully exploit the sensor’s higher operating speed measuring functions ´Maximum
value´, ´Minimum value´, and ´Average value´ are available. These values are ac-
quired at 200 mops (measuring operations per second); they are formed and output in
all measured value scans (continuous or cyclic) in synchrony with the scan of the 1st
sensor channel (normally measured pressure).
20.3 Technical data
Measuring ranges
Pressure depending on type (see data sheet)
Resolution 0.002 % full-scale (FS)
Accuracy ±0.05 % full-scale (FS) (+10 to +40 °C),
±0.1 % full-scale (FS) (-10 to +80 °C)
Temperature -40 to +120 °C
Resolution 0.01 K
Sensor’s measuring rate
200 mops (measuring operations per second)
Setting time
0.6 seconds
Delay after sleep mode
1 second
Refresh rate
0.005 seconds for all channels
Connector colors
2 colors, light gray and dark gray, red lever
Standard baud rate
115.2 kbaud (freely selectable from 1200 baud up to 921
kbaud)
Supply voltage
6 to 13 VDC
Current consumption
approx. 11 mA
- 43 -