12.2.2 Emissivity and transmittance
The emissivity of a measured object (see Manual, 3.1.5) is important in ensuring reliable
measured results; this material-dependent variable (factory default 0.95) can be set either in
the sensor menu or in the normal V6 sensor programming. If the latter method is used and
an infra-red sensor is connected, ´gain correction´ will be replaced by ´emission´, so that
emissivity can be programmed here in the normal way.
If a protective window is being used, these calculations may also have to take account of
transmittance (factory default 1.00). However, this quantity can only be set in the sensor
menu.
12.3 Technical data
Operative range
Probe head
-10 to +120 °C
Measuring ranges
Temperature
-40.0 to +600.0 °C
Accuracy
±1 % of measured value or ±1 K
Temperature coefficient ±0.5 K / K or ±0.05 % / K
Refresh rate
0.25 seconds for all channels
Connector colors
2 colors, light gray and dark gray, red lever
Standard baud rate
115.2 kbaud (freely selectable from 9600 baud up to 921
kbaud)
Supply voltage
6 to 13 VDC
Current consumption
4 mA
Sleep mode on the device
Possible (for extensions a 1s delay is necessary)