8.2 Device Address and Networking
All ALMEMO
®
instruments can be very easily networked to centrally acquire the
measured values of several instruments that are located at different places
(see manual 5.3). For communicating with networked devices it is mandatory
that each device has its own address as only one device is allowed to respond
to each command. Therefore, before any network operation it is necessary
that all connected devices are set to different device numbers. This is
performed with the rotary switch in position OUTP. CYCLE. The key
FUNCTION
is used to select the function DEVICE ADDRESS ´
A
´ and the currently set
device number is displayed, which is usually factory-set to 00. It can then be
modified by the normal data entry (see 5.3).
Function:
OUTP. CYCLE
FUNCTION
0 1
A
Example:
Address 01
Only successive numbers between 01 and 99 should be entered for network
operation so that the device 00 cannot be falsely addressed in case of a power
supply failure.
8.3 Manual Data Output
For the output of data to a printer the output channel must be set to ´
U
´ by
using the key
CH
within the function PRINT CYCLE (see 7.3.1). The output
format is not relevant for the manual data output, with the exception of the
memory. All function values that have been selected by using the rotary switch
and, possibly, with the key
FUNCTION
, can be printed out with the next print
output by using the key
OUTPUT
.
KEY
... Manual Data Output:
FUNCTION
OUTPUT
Switch
Func Key Print Output
MEAS. VAL.
12:34:00 01: +0023.5 °C
MAX VALUE
MAXIMUM: 01: +0020.0 °C
MIN VALUE
MINIMUM: 01: -0010.0 °C
AVERAGE V.
AVERAGE VAL: 01: +0017.8 °C
AVERAGE V.
all meas. values
AV
C
F
F, F
MS MEAS.VAL. MAXIMUM MINIMUM AVG COUNT
01: +0023.0 +0025.0 +0019.0 +0022.0 99999
man. 6.4.4
NUMBER
NUMBER: 00-123
MEMORY
MEMORY: - - - -
see 7.4.2
MEMORY
MF
F
MEMORY: S0501.3 F0324.6 A
42
ALMEMO
3290-8
Analogue Output