8. DIGITAL DATA OUTPUT
The entire programming of the sensors and the instrument, as well as all
measured values, can be provided as output to a printer or computer via serial
interface. The interface modules and the connection to the instruments are
described in the manual section 5.2. Other modules for networking the
instruments follow in the manual section 5.3.
8.1 Baud Rate, Data Format
All interface modules are factory-set and programmed to 9600 baud. To avoid
unnecessary problems when networking several devices the baud rate should
not be modified but the computer or printer should be set up accordingly. If this
is not possible, the values 150, 300, 600, 1200, 2400, 4800, 9600 or 57600bd
can be entered via keyboard if the rotary switch is in position BAUD RATE. The
input is started with the key
ENTER
. The display will start to flash and can be
modified by using the keys
and . When the desired transmission rate has
been selected the programming can be terminated by operating the key
ENTER
once again. The baud rate setting will be stored in the EEPROM of the
interface module and will then be valid for use with all other ALMEMO
®
devices.
Function:
BAUD RATE
9 6 0 0
BR
Example:
9600 bd
Data format:
unchangeable 8 data bits, no parity, 1 stop bit
ALMEMO
3290-8
41
Digital Data Output
Summary of Contents for ALMEMO 3290-8 V5
Page 48: ...48 ALMEMO 3290 8 Appendix...