![Ahlborn ALMEMO 3290-8 V5 Operating Instructions Manual Download Page 23](http://html1.mh-extra.com/html/ahlborn/almemo-3290-8-v5/almemo-3290-8-v5_operating-instructions-manual_2870446023.webp)
6.3 Changing the Dimension
Each measuring channel allows to replace the standard dimension of the
measuring range by any other dimension that has two digits (see manual
6.3.5). In addition to all capital and normal letters, the characters
,
Ω
, %, [,
,
], *, -, =, ~ and spaces (_) are available. The dimension is indicated by two
16-segment characters that are indicated following the measuring and
programming values.
The
change of the dimension
can be performed within the function RANGE
by pressing the key
ENTER
and
ß
. The first character of the dimension will flash
in the display. It can then be changed by using the keys
×
and
Ø
. When the
first character is selected the key
ENTER
should be pressed once again and the
same procedure will be performed for the second character. When the desired
dimension has been set the programming can be completed by the key
ENTER
.
RANGE
,
,
... ,
,
... ,
ENTER
ENTER
1st dim change
2nd dim change
End
When the dimension °F is entered a temperature value in degrees
Celsius will be converted into degrees Fahrenheit.
The cold junction compensation can be switched off by using the
character .
The dimension
ms
is indicated on the display as
m
/
s
, and
mh
as
m
3
/
h
.
6.4 Limit Values
Two limit values (MAX and MIN) can be programmed for each measuring
channel. The exceeding of the limit values is handled as a fault, similar to the
exceeding of the measuring range limits and sensor breakage. The arrow
ALARM will appear in the display and the alarm relays will respond and the
alarm values will be provided as output during the measuring cycle (see
manual 6.3.9). An exceeding can also be used to start or stop a measuring
point scan (see 7.3.6).
Function:
Limit value Max
LV: MAX
Limit value Min
LV: MIN
Programming
Input according to 5.3
1: 1 2 3.0
°C
Switch-off
,
ENTER
Clr
1: - - - -
°C
ALMEMO
3290-8
23
Sensor Programming
Summary of Contents for ALMEMO 3290-8 V5
Page 48: ...48 ALMEMO 3290 8 Appendix...