7.1.2 Memory for Peak Values
From the acquired measured values of each measuring point the highest and
lowest value is determined and stored. For indicating the peak values the
function MAX. VAL. or MIN. VAL. must be selected with the rotary switch and
the desired channel must be set.
Function:
Peak values:
MAX. VAL.
MIN. VAL.
Clear:
,
The peak values are cleared if a total clearing (see 7) or change of the range
(see 6.2) is carried out. If the cleared channel is the selected measuring
channel, the measured value will be indicated immediately after the clearing.
7.1.3 Setting Measured Value to Zero, Zero Point Correction
Setting the Measured Value to Zero
The user can zero the measured value at certain locations or at certain times in
order to check the deviation from this reference value. The indicated measured
value is, by the following key combination, stored as base value and, as a
result, set to zero.
Function:
MEAS. VALUE
Zero setting by keys:
,
ENTER
Clr
Please note that this function is only available if the locking code is
set below 5 (see 6.7).
The arrow
CORR
. appears in the display as long as the deviation
from the base value is indicated, but not the actual measured
value.
The base value must be cleared in order to obtain the actual measured value
(see 6.6). For this purpose the rotary switch must be set to the function BASE
and the base value must be cleared with the keys
ENTER
,
Clr
.
Function:
BASE
Clear base value:
,
ENTER
Clr
ENTER
Clr
28
ALMEMO
3290-8
Continuous Measurement of a Measuring Point