2. INITIAL OPERATION
1. Connect
transducers
to the sockets M0 to M8 (5), see 4.
2. Ensure
power supply
by mains adapter connected to socket (7), see 3.1
3.
For switching on
set the push button (4) so the control lamp is on, s. 3.4.
4.
For displaying
the measured values,
select function MEAS. VALUE by using the rotary switch (3),
use key
CH
to select the measuring channel, read meas. value, see 7.1.
5.
For storing
the measured values:
Use function MEMORY and keys
ENTER
,
Clr
to clear the memory, s. 7.4.2.
Use MEAS. CYCLE and key
CH
to activate the memory, see 7.3.2.
Single storing by using the key
MANUAL
, see 7.2.
Enter measuring cycle for cyclic storing, see 7.3.2.
Enter time and date, as required, see 7.3.4.
Enter time and date of start or end of a measurement as required, see 7.3.5.
Use key
START/STOP
to start and stop a cyclic storing, see 7.3.
Output of memory data to printer or computer
Connect peripheral device via interface cable to socket A1, see manual 5.2.
Set 9600 bd, 8 data bits, 1 stop bit, no parity at peripheral device.
Use key
CH
in function OUTP. CYCLE to set the output channel ´
U
´ and,
possibly, output format columns ´
nU
´ or spreadsheet/table ´
tU
´, see 7.3.1.
Use key
OUTPUT
within function MEMORY to output meas.values, see 7.4.2.
6.
Cyclic output of measured values
to printer or computer
Connect peripheral device via interface cable to socket A1, see manual 5.2.
Set 9600 bd, 8 data bits, 1 stop bit, no parity at peripheral device.
Enter time and date, as required, see 7.3.4.
Program the print cycle within function OUTP. CYCLE,
use key
CH
to set the output channel ´
U
´ and, if required,
the output format columns ´
nU
´ or spreadsheet/table ´
tU
´, see 7.3.1.
Use key
START/STOP
to start and stop the cyclic meas. point scan, see 7.3.
7.
Monitoring of limit values
Enter limit values, see 6.4.
Program measuring cycle, see 7.3.2.
Connect alarm device with alarm module to socket A2, see man. 5.1.2/5.1.3.
For an alarm print use key
CH
within function OUTP. CYCLE to activate
the output channel ´ U´, see 7.3.1.
Use key
START/STOP
to start and stop the cyclic meas. point scan, see 7.3.
8.
Evaluation of the measurement
Display max and min values within function MAX.VAL. or MIN.VAL.,
see 7.1.2.
12
ALMEMO
3290-8
Initial Operation
Summary of Contents for ALMEMO 3290-8 V5
Page 48: ...48 ALMEMO 3290 8 Appendix...