7.3.3 Conversion Rate, Continuous Measuring Point Scan
If required the conversion rate can be increased from 2.5 to 10M/sec (see
manual 6.5, 6.5.4). The rotary switch must be moved to position MEAS.
CYCLE and the additional function CONVERSION RATE ´
CR
´ must be selected
by using the key F and must be set by using the keys
ENTER
,
▲
▼
,
ENTER
.
At the same time, the
continuous measuring point scan
(see man. 6.5.1.3)
can be set with coding ´
C
´, i.e. not only the selected measuring point but all
active measuring channels are scanned successively without interruption. The
storage with the conversion rate (coding ´
S
´) can be activated with key CH
▲
,
the output of the measured values ´
U
´ can only be activated via the interface.
Function:
Conversion Rate
(
CR
)
MEAS. CYCLE
Key:
Example:
10M/s, continuous, with saving
Change conversion rate:
,
... ,
Memory on/off:
ENTER
ENTER
CH
▲
7.3.4 Time and Date
The ALMEMO
®
2290-8 is equipped with a real time clock with date function for
recording the measuring time. It has a lithium battery so the time and date are
maintained after a switch-off.
Function Time:
TIME
1 2:3 4:5 6
TM
The time is programmed in the format
hh:mm:ss
(see 5.3).
Stopping the clock and setting it to zero can be performed by using the keys
ENTER
,
CLEAR
.
The clock can be started in any switch position by using the key
START/STOP
.
Function Date:
DATE
0 1:0 5:9 9
DA
Example:
date 1st May 1999
Enter the date in the format
dd.mm.yy
(see 5.3). The year number can also be
provided with 4 digits via interface (see manual 6.10.13).
Clear the date by using the keys
ENTER
,
CLEAR
.
F
1
0 C S
CR
36
ALMEMO
®
2290-8
Cyclic Measuring Point Scan