![Ahlborn ALMEMO 204 Operating Instructions Manual Download Page 26](http://html1.mh-extra.com/html/ahlborn/almemo-204/almemo-204_operating-instructions-manual_2870457026.webp)
11. Measuring menus
11. MEASURING MENUS
The menu
Meas. points list
provides not only the sensor display (listing all
measured values for a particular sensor plus any appropriate compensation
values) but also a clear overview of all measuring channels together with the
most important data on each (s. 11.3). To output measured values to an inter-
face or memory at a certain scanning rate and output cycle, you can select
user menu
U2 Data logger
. If these do not completely meet your require-
ments, you can also assemble your own user menu from a range of over 50
functions (s. 13). Each measuring menu can, by means of function menus, also
be assigned various functions (s. 14).
11.1 Sensor display menu
Via the sensor list you can access the intelligent
Sensor display
menu. In the first line you will see
the measured value (up to seven digits in length,
large format), the measuring point, and the units
(up to six characters in length, small format). Be-
low this appear the measuring point designation
(up to 20 characters in length) and certain symbols
for checking the measured value status (s. 9.2). Below this, depending on the
quantity and range, it lists all functions of importance for this measured value
(e.g. compensation values) plus any further measuring channels assigned to
the sensor in question.
Additional measuring functions can be implemented by means of function
menus (s. 14). Symbol
<M>
in the middle of the softkey bar indicates that the
measuring point can be selected by pressing keys ▲ / ▼.
11.1.1 Measuring channel selection
With key
all active measuring channels can be selected one after the
other and the latest measured value of each can be displayed. With key
the previous channel is displayed. When a particular measuring channel is se-
lected the associated input channel is also selected at the same time.
When performing this selection it should be noted that with this V7
device the channel numbering system has been changed; the chan-
nels are now numbered per sensor.
To increment the measuring channel:
To decrement the measuring channel:
11.2 Measured value correction and compensation
To achieve maximum measuring accuracy zero-point correction can be per-
formed for sensors as early as the
Sensor display
menu. Universal two-point
adjustment can be performed for all sensors via the function menus
Two-point
adjustment
(with two actual values and two setpoint values) and
Scaling
(s.
26 ALMEMO
®
204
Summary of Contents for ALMEMO 204
Page 72: ...20 Annex 72 ALMEMO 204...