15. Sensor programming
Once the scaling values have been programmed and the actual measured
value thus modified, the correction arrow appears indicating the measured
value status
º
(s. 9.2).
15.8 Changing the units
For each measuring channel the default units for the quantity and measuring
range can be replaced with any other units, two characters for standard sen-
sors, up to six characters for D7 sensors (see Manual 6.3.5). All upper-case
and lower-case letters, special characters °, Ω, %, [, ], *, -, =, ~, and space (_)
can be used. The units are shown after the measured value or programming
value in question.
To change the units use the function:
1 Range, Units: DIGI
°C
If
°F
is entered as units the temperature value will be converted auto-
matically from degrees Celsius to degrees Fahrenheit. Entering the
appropriate two characters will automatically generate the following
units:
for
m¡
enter
ms
, for
¦¢
enter
mh
, for
¤¥
enter
Wm
, and for
g£
enter
gk
..
15.9 Selecting the measuring range
With this device the individual measuring ranges for D6 and D7 sensors can
only be modified via the 'sensor configuration' menu (s. 15.10). This section
describes which function channels can also be used. Please note that locking
is always completely canceled, i.e. the locking level is reset to zero (s. 15.4).
To activate a new measuring channel key
<MALL>
should first be pressed; this
accesses all selectable channels; then select the required input channel, and
then enter the measuring range (s. 15.1). When the new measuring range en-
try is confirmed all programming values for that input channel will be deleted.
Function, select measuring range
Range, Units:
DIGI
°C
To access all possible measuring channels
<MALL>
To deactivate a channel
<OFF>
To activate a channel
<ON>
Programming the range is as for data input 9.5
PROG
,
▲
,
▲
,
PROG
In the input window all abbreviations appear one
after the other, as listed in the following table
Range:
DIGI
There is also a help window for identifying the
sensor in question.
48 ALMEMO
®
204
xxxx
Average values over
time M(t)