18
- T
he “Emergency Emptying System” cable circuit (Accessory) is damaged/disconnected.
-
The “Emergency Stop Jumper” (Accessory) is missing, defective or incorrect mounted.
Press
“Stop” button, check equipment, release “Emergency Stop Button” and retry.
"Sensor-Error"
- The pressure sensor is sending improbable values to the computer.
- Try switching the PCI off and on.
If the error persists then contact AHCON or your AHCON product dealer/distributor.
"CRC Error"
-
The PCI software has detected a “Checksum” error. Try switching the PCI off and on
again.
If the “CRC Error” message persists the PCI is defective.
Please contact AHCON or your AHCON dealer/distributor for repair.
Maintenance and Repair
Use only trained personnel for maintenance and repair of this equipment or c
ontact
AHCON or your AHCON product dealer/distributor.
Regular cleaning maintenance consists solely of cleaning the PCI cover and tubing with a
lightly damp cloth with a mild solution of general purpose detergent liquid.
Contact your AHCON dealer or Flex1one A/S (AHCON) for any repair.
Accessories
Article number:
Description:
0502 - 900 52 007
Wall Mounting Bracket, PCI 900/1000/1200.
0502 - 900 52 009
Wall Mounting Bracket, small, PCI 900/1200.
0502 - 100 20 010 Door switch (2 pcs.) with cable.
0504 - 950 13 006 Emergency Emptying System (Emergency Stop w/10 m cable).
0350 - 4000 0130
Signal Lamp, orange (available in other colors).
0350 - 4000 0135
Combined light and sound-giving device (color options).
0502 - 970 00 002 Door Jumper (Cable lice).
0502 - 970 00 003 Emergency Jumper (Cable lice).