56

HOW TO OPERATE WITH ADAPTER 

WARNINGS

 Only use an external adaptor that is fulfilled in compliance with 

the requirements of IEC60601–1:2005. Please choose AC 5V / 1A adapter, and 

the connection jack should match a plug with outer diameter of 3.5 mm, inte-

rior diameter of 1.35 mm, 9.5 mm in length and positive polarity                  .

 

The adapter does not charge batteries. Please remove the batteries before 

using the adapter.

 

As  with  all  electronic  devices,  it  is  recommended  that  keep  the  product 

unplugged when not in use.

 

+

-

1.  Set up the medication cup

 

Align the three rubs on the back 

of the medication cup with three 

trenches of the main unit to insert 

the medication cup into the main 

unit.

2. Fill in the prescribed medication 

as illustrated

Open the medication cup cover.

 

Fill  the  mediction  cup  with 

prescribed medication. 

The maxi-

mum capacity is 6 ml.

 

Close  the  medication  cup  cover 

properly.

Summary of Contents for Minimill N7

Page 1: ...Minimill MESH NEBULIZER AGU N7 DE BEDIENUNGSANLEITUNG FR MODE D EMPLOI EN INSTRUCTION MANUAL RU BG agu baby com...

Page 2: ...erheitshinweise 6 5 Zubeh rteile und Teilbezeichnung 7 6 Gebrauchsvorschrift f r Batterien 9 7 Gebrauchsvorschrift f r den Netzadapter 12 8 Reinigung und Desinfektion 14 9 Fehlerbehebung 16 10 Technis...

Page 3: ...lle der Einwegbatterien die wieder aufladbaren AA NiMH Akkus optionales Zubeh rteil im Pack nicht enthalten zu kaufen Der Vernebler AGU N7 kann auch eine Hilfsspannungsquelle in Form eines Netzadapter...

Page 4: ...magnetische Vertr glichkeit der Betriebsmittel ber die elektromagnetische Vertr glichkeit der Betriebsmittel den Hauptanforderungen TR ZU 004 2011 ber Sicherheit von Niederspannungsanlagen WEEE Richt...

Page 5: ...s und den Inhalt dieser Bedienungsanleitung zu verstehen Patienten Dieses Ger t ist f r alle Altersgruppen der Patienten geeignet au er f r die Patienten die bewusstlos sind nicht spontan atmen oder a...

Page 6: ...e Verwendung verursacht werden Teilen Sie Ihren Vernebler nicht mit anderen Es ist beabsichtigt dass der Vernebler von einem einzigen Benutzer verwendet wird Beim Verwenden von mehreren Personen best...

Page 7: ...e AGU MC7 Mundst ck Medikamentenschale AGU MC7 Mundst ck 4 Bedienung sanleitung 3 Kinder maske Kindermaske Um Ersatzteile oder Zubeh r auszutauschen wenden Sie sich bitte an Ihren lokalen H ndler oder...

Page 8: ...euchtanzeige Gr nes Licht Dauerbetrieb Oranges Licht konstant oder blinkt schwache Leistung St rung Keine Leuchtanzeige Fehlfunktion Abdeckung vom Batteriefach Netzkabelanschluss Abdeckung des Netzkab...

Page 9: ...um Batterien wieder aufzuladen Reinigen Sie alle Teile Ihres tragbaren Verneblers vor Gebrauch nach jedem Gebrauch und nach l ngerer Lagerung 1 Legen Sie die Batterien ein Drehen Sie das Ger t um Dr c...

Page 10: ...ordentlich zu 4 Inhalation Die Ein Aus Taste dr cken um das Ger t einzuschalten Leuchtanzeige leuchtet gr n Nehmen Sie das Mundst ck zwischen Ihre Z hne und halten Sie es mit den Lippen fest Wenn Sie...

Page 11: ...ten Sie in diesem Fall die Stromversorgung aus und nehmen Sie die L sung mit einem Gazetuch oder mit einem fusselfreien Stoff auf Batterielebensdauer und Batteriewechsel Die Batterielebensdauer h ngt...

Page 12: ...dem Ger t bevor Sie den Adapter verwenden Wie bei allen elektronischen Ger ten empfiehlt es sich diesen Vernebler bei Nichtgebrauch aus der Steckdose zu ziehen 1 Stellen Sie die Medikamentens chale a...

Page 13: ...dr cken um das Ger t einzuschalten Leuchtanzeigeleuchtetgr n Nehmen Sie das Mundst ck zwischen Ihre Z hne und halten Sie es mit den Lippen fest Wenn Sie eine Maske verwenden legen Sie die Maske ber Mu...

Page 14: ...h tteln Sie bersch ssiges Wasser ab und lassen Sie die Zubeh rteile auf einem sauberen trockenen Handtuch vollst ndig lufttrocknen Falls n tig benutzen Sie ein sauberes Handtuch um nasse Flecken von d...

Page 15: ...Stellen Sie sicher dass alle Teile vor Lagerung und neuem Gebrauch sauber und trocken sind WARNUNGEN Alkohol ist leicht entz ndlich Verwenden Sie keinen Alkohol in der N he von offenem Feuer oder Rauc...

Page 16: ...e leuchtet nicht und kein Nebel kommt heraus Die Batterien wurden in die falsche Richtung eingesteckt Markierung und auf der Abdeckung des Batteriefaches beachten um die Batterien gem der angegebenen...

Page 17: ...dikamentenschale ist nicht richtig installiert Siehe die Anleitung ber die Medikamentenschale um diese von neuem zu installieren Membran ffnungen sind verstopft Reinigen Sie die Medikamentenschale sie...

Page 18: ...AGU N7 Funktionsweise Aktive Schwingungsmembran Technologie Stromversorgung AA Alkalische Batterie x 2 im Pack enthalten AA NiMH wieder aufladbare Batterie x 2 im Pack nicht enthalten Netz Adapter im...

Page 19: ...oder hochviskose Arzneimittel anwendbar Weitere Informationen erhalten Sie beim Hersteller der entsprechenden Arzneimittel Hinweis Dieses Ger t und sein Zubeh r entsprechen der europ ischen Norm EN 13...

Page 20: ...er die Anwender des Membran Verneblers AGU N7 sollten sicherstellen dass er in solcher Umgebung eingesetzt wird Immunit ts Test IEC 60601 Pr fniveau Compliance Niveau Elektromagnetische Umgebungs Bedi...

Page 21: ...n IEC 61000 4 11 5 UT 95 Dip in UT f r 0 5 Zyklen 40 UT 60 Dip in UT f r 5 Zyklen 70 UT 30 Dip in UT f r 25 Zyklen 5 UT 95 Dip in UT f r 5 Sek 5 UT 95 Dip in UT f r 0 5 Zyklen 40 UT 60 Dip in UT f r 5...

Page 22: ...n weniger als 3 V m betragen Immunit ts Test IEC 60601 Pr fniveau Comp liance Niveau Elektromagnetische Umgebungs Bedingungen Geleitete HF IEC 61000 4 6 Ausgestrahlte HF IEC 61000 4 3 3 V m 150 KHz bi...

Page 23: ...arantie gegen Defekte in Material und Verarbeitung f r einen Zeitraum von 2 Jahren ab dem Kaufdatum Die Garantie gilt nicht f r den Batteriebetrieb Folgen Sie den Anwei sungen um die sichere und langf...

Page 24: ...es 29 6 Piles Conditions d utilisation de l appareil 31 7 Adaptateur r seau Conditions d utilisation de l appareil 34 8 Nettoyage et d sinfection 36 9 Dysfonctionnements possibles et moyens de correc...

Page 25: ...au lieu des piles usage unique Le n buliseur AGU N7 peut avoir une alimentation lectrique suppl mentaire sous la forme d un adaptateur r seau non inclus dans le kit L adaptateur vous permet d conomise...

Page 26: ...de lot Cet appareil est conforme aux exigences de la directive sur les dispositifs m dicaux 93 42 EEC Repr sentant autoris dans l UE L appareil est conforme aux exigences principales du R glement tech...

Page 27: ...ositif Patients potentiels Ce dispositif est destin aux patients qui peuvent respirer par eux m mes en conscience en l absence de signe d d me pulmonaire Il n y a pas de restriction d ge pour l utilis...

Page 28: ...sabilit pour tout dommage de dispositif r sultant d une mauvaise utilisation ou d un mauvais entretien Ne partagez pas votre n buliseur avec d autres personnes Le n buliseur est uniquement destin un u...

Page 29: ...dommages veuillez contacter le magasin ou la pharmacie o l appareil a t achet PI CES DE RECHANGE ET ACCESSOIRES 1 Unit principale 2 Brumisateur AGU MC7 Embout Brumisateur AGU MC7 Embout buccal 4 Manu...

Page 30: ...L indicateur fixe ou clignotant orange piles faibles dysfonctionnement L indicateur ne s allume pas dysfonctionnement Couvercle de la prise c ble r seau Prise c ble r seau Couvercle de la prise c ble...

Page 31: ...les batteries Nettoyez toutes les parties du n buliseur avant la premi re et chaque utilisation suivante ainsi qu apr s une longue pause 1 Insertion des piles Retournez l appareil Faites glisser le co...

Page 32: ...e bouton ON OFF sur la partie sup rieure d appareil et mettez le en marche l indicateur deviendra vert Placez l embout buccal entre vos dents et enroulez troitement vos l vres autour Si vous utilisez...

Page 33: ...uliseur peut s arr ter Dans cette situation d branchez le du r seau et pongez la solution avec un morceau de gaze ou un chiffon non pelucheux La dur e de vie et de remplacement des piles La dur e de v...

Page 34: ...ne charge pas les piles et les accumulateurs Avant d utiliser l adaptateur secteur retirez les piles de l appareil Comme tous les appareils lectroniques cet appareil doit tre d connect du r seau pend...

Page 35: ...tie sup rieure de l appareil et mettez le en marche l indicateur deviendravert Placez l embout buccal entre vos dents et enroulez troitement vos l vres autour Si vous utilisez le masque mettez le sur...

Page 36: ...ropre Assurez vous que toutes les parties nettoy es de l appareil sont compl tements s ches avant de les mettre en stockage et les utiliser la prochaine fois ATTENTION Veuillez ce que le compartiment...

Page 37: ...port et stockage de l appareil Gardez l appareil et le pistolet dans un endroit propre et sec ATTENTION Ne laissez pas et ne transportez pas l appareil qui contient le reste de la liquide dans le pist...

Page 38: ...ns le mode ON l indicateur n est pas allum et l a rosol ne sort pas La polarit des piles n est pas respect e Installez les piles en respectant la polarit et Les piles sont d charg es Changez les piles...

Page 39: ...sans piles Conditions d exploitation de 10 C 40 C humidit relative maximum 30 85 800 1060 kPa Conditions de transport et de stockage de 20 C 70 C humidit relative maximum 20 75 800 1060 kPa Dur e de...

Page 40: ...s ont t effectu es en utilisant une solution de NaCl l aide d un testeur NGI Impacter de la G n ration suivante Ce sch ma ne peut pas tre appliqu pour des suspensions et des solutions m dicinales de h...

Page 41: ...lectromagn tique tel que d crit ci dessous Les acheteurs ou les utilisateurs de l appareil AGU N7 doivent veiller son utilisation dans l environnement correspondant Essai d mission Conformit Directive...

Page 42: ...Soubresauts DME 61000 4 5 1 kV entre les lignes 2 kV entre la ligne et la terre 1 kV r gime diff rentiel n a pas de correspondances La qualit de l alimentation dans le r seau doit tre conforme aux no...

Page 43: ...en watts W indiqu e par le fabricant et d est la distance minimale recommand e en m tres m Les valeurs de r sistances mesur es du champ lectromagn tique produit par l metteur radio fixe doivent tre m...

Page 44: ...5 GHz d 2 3 P 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73 1 1 2 1 2 2 3 10 3 8 3 8 7 3 100 12 12 23 Pour les metteurs avec la puissance nominale maximale de rayonnement diff rente la distance minimale reco...

Page 45: ...ne fonctionne pas normalement veuillez vous adresser au vendeur ATTENTION N essayez pas de r parer le n buliseur vous m me cela entra ne l annulation de la garantie La garantie sera annul e si on con...

Page 46: ...ed use 49 4 Safety precautions 50 5 Package contents and overview 51 6 How to operate with batteries 53 7 How to operate with adapter 56 8 Cleaning and disinfection 58 9 Troubleshooting 60 10 Specific...

Page 47: ...e also recommend you to buy AA NiMH recharge able batteries optional item not included instead of disposable batteries Mesh Nebulizer AGU N7 has extra power option it can be driven by both two AA batt...

Page 48: ...on Electromagnetic Compatibility of Technical Products The unit meets the main requirements of Customs Union Technical Regulation 004 2011 On the safety of low voltage equip ment concerning the safet...

Page 49: ...is device and the content of instruction manual Intended patients This device design for all ages of patients Except for the patients who are uncon scious not breathing spontaneously or having pulmona...

Page 50: ...here is a risk of spreading infectious illness Clean all parts of your nebulizer before use after each use and after extend ed storage Adult supervision is required when this device is used by childre...

Page 51: ...ny damaged parts before you use this device In the case of missing parts malfunction or damage please contact the store where you purchased from or the nearest dealer OPTIONAL PARTS AND ACCESSORIES 1...

Page 52: ...ing Operation Orange light on flash Low power Malfunction No light Malfunction Bottom view Power Cable Jack Medication Cup Cover Medication Cup Mesh Mouthpiece ON OFF Button Electrode Press ON OFF But...

Page 53: ...le nebulizer before use after each use and after extended storage 1 Insert the batteries Turn the unit upside down Push back and pull up the battery cover to open it Insert two AA alkaline NiMH rechar...

Page 54: ...p cover properly 4 Inhalation Press ON OFF button on the top to turn on the device light Indica tor turns on in green Place the mouthpiece between your teeth with your lips firmly sealed around the mo...

Page 55: ...rn off the power and absorb the solution by gauze or lint free towel Battery life and battery replacement Battery life is usually depending on the capacity and condition of the batteries In general th...

Page 56: ...charge batteries Please remove the batteries before using the adapter As with all electronic devices it is recommended that keep the product unplugged when not in use 1 Set up the medication cup Alig...

Page 57: ...ss ON OFF button on the top to turn on the device light Indi cator turns on in green Place the mouthpiece between your teeth with your lips firmly sealed around the mouthpiece If you use a mask place...

Page 58: ...gauze or clean towel to wipe off stains on the main unit if necessary Make sure that all cleaned parts are completely dry before you store them or use them next time WARNINGS Keep the battery compart...

Page 59: ...r alcohol The following items are able to be disinfected by boiled water or alcohol CARRYING AND STORING Put on the mouthpiece cover store the device and medication cup in a dry and clean environment...

Page 60: ...ation even if power switch is on power indicator does not light and no mist comes out The batteries are inserted in the wrong direction Follow and mark on the battery cover to re insert the batteries...

Page 61: ...humidity 10 40 C 30 85 RH 800 1060 hPa Storage and delivery condition 20 70 C 20 75 RH 800 1060 hPa Durable period Main unit Medication cup mask for children 5 years not less than 6 months with proper...

Page 62: ...information can be obtained from the relevant medicine manufacturer Note ThisdeviceanditsaccessoriescomplywithEuropeanstandardEN13544 1 2007 A1 2009 Particle Size MMAD 4 5 m MMAD Mass Median Aerodyna...

Page 63: ...ded for use in the electromagnetic environ ment specified below The customer or the user of the Mesh nebulizer membrane AGU N7 should assure that it is used in such an environment Emission test Compli...

Page 64: ...pical commercial or hospital environment Surge IEC 61000 4 5 1 kV line s to line s 2 kV line s to earth 1 kV differential mode Not applicable Main power quality should be that of a typical commercial...

Page 65: ...as determined by an electromagnetic site surveya should be less than the compliance level in each frequency rangeb Interference may occur in the vicinity of equipment marked with the following symbol...

Page 66: ...at a maximum output power not listed above the recommend ed separation distance d in metres m can be estimated using the equation applica ble to the frequency of the transmitter where P is the maximu...

Page 67: ...ce does not work properly contact the authorized seller WARNINGS Do not attempt to repair the nebulizer by yourself it will lead to a loss of warranty Changes made to the nebulizer without the manufac...

Page 68: ...68 1 69 2 70 3 71 4 72 5 73 6 75 7 78 8 80 9 82 10 83 11 84 12 89 AGU N7 RU...

Page 69: ...69 RU AGU N7 AGU N7 AGU N7 AA AGU N7 AGU N7 Active Vibrating Mesh AGU N7...

Page 70: ...70 II IEC 60601 1 BF IEC 60601 1 12 5 15 ON OFF DC 93 42 EEC 020 2011 004 2011 WEEE...

Page 71: ...71 RU M 10 C 40 C 30 85...

Page 72: ...72...

Page 73: ...73 RU 1 2 AGU MC7 AGU MC7 4 3...

Page 74: ...74...

Page 75: ...75 RU AGU N7 AA NiMH IEC 62133 1 AA NiMH...

Page 76: ...76 3 6 4 2...

Page 77: ...77 RU 5 AGU N7 4 AA 20 AA AA NiMH...

Page 78: ...78 IEC60601 1 2005 DC 5V 1A 3 5 1 35 9 5 1 2 6...

Page 79: ...79 RU 3 4 5...

Page 80: ...80 1 2 1 5...

Page 81: ...81 RU 2 5...

Page 82: ...82...

Page 83: ...3 RU AGU N7 AGU N7 MBPN002 Active Vibrating Mesh AA x 2 AA NiMH x 2 1 5 117 15 0 35 MMAD 4 5 6 76 5 x 41 x 73 74 10 C 40 C 30 85 800 1060 20 C 70 C 20 75 800 1060 5 6 2 3 IEC60601 1 2005 IEC62133 23 C...

Page 84: ...84 MMAD 4 5 MMAD 0 35 6 23 1 0 20 40 60 80 100 0 1 1 10 100 0 20 40 60 NaCl NGI EN 13544 1 2007 A1 2009...

Page 85: ...85 RU EMC EN60601 1 2 AGU N7 MBPN002 EN60601 1 2 2007 AGU N7 MBPN002 AGU N7 AGU N7 CISPR 11 1 AGU N7 CISPR 11 AGU N7 61000 3 2 61000 3 3...

Page 86: ...AGU N7 AGU N7 60601 61000 4 2 6 8 6 8 30 61000 4 4 2 1 2 61000 4 5 1 2 1 61000 4 11 5 UT UT 95 40 UT UT 60 5 70 UT UT 30 25 5 UT UT 95 5 c 5 UT UT 95 40 UT UT 60 5 70 UT UT 30 25 5 UT UT 95 5 c AGU N...

Page 87: ...87 RU 50 60 61000 4 8 3 A 3 A AGU N7 61000 4 6 61000 4 3 3 150 80 3 80 2 5 3 3 AGU N7 d 1 2 P d 1 2 P 80 800 d 2 3 P 800 2 5 P d a b...

Page 88: ...T 1 80 800 2 a FM AGU N7 AGU N7 AGU N7 b 150 80 3 AGU N7 AGU N7 AGU N7 AGU N7 150 80 d 1 2 P 80 800 d 1 2 P 800 2 5 d 2 3 P 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73 1 1 2 1 2 2 3 10 3 8 3 8 7 3 100 12 1...

Page 89: ...89 RU d P 1 80 800 2 2 agu baby com...

Page 90: ...90 1 91 2 92 3 93 4 94 5 95 6 97 7 100 8 102 9 104 10 105 11 106 12 111 AGU N7 BG...

Page 91: ...91 BG AGU N7 AGU N7 AGU N7 AGU N7 AGU N7 Active Vibrating Mesh AGU N7...

Page 92: ...92 II IEC 60601 1 BF IEC 60601 1 12 5 15 ON OFF DC 93 42 EEC 020 2011 004 2011 WEEE...

Page 93: ...93 BG 10 40 30 85...

Page 94: ...94...

Page 95: ...95 BG 1 2 AGU MC7 AGU MC7 4 3...

Page 96: ...96...

Page 97: ...97 BG AGU N7 AA NiMH IEC 62133 1 NiMH...

Page 98: ...98 3 6 4 2...

Page 99: ...99 BG 5 AGU N7 4 20 NiMH...

Page 100: ...100 IEC 60601 1 2005 AC 5V 1A 3 5 mm 1 35 9 5 1 2 6...

Page 101: ...101 BG 3 4 5...

Page 102: ...102 1 2 1 5...

Page 103: ...103 BG 2 75 1 5...

Page 104: ...104...

Page 105: ...N7 MBPN002 Active Vibrating Mesh AA x 2 AA NiMH x 2 1 5 W 117 kXz 15 0 35 ml min MMAD 4 5 m 6 ml 76 5 x 41 x 73 mm 74 g 10 C 40 C 30 85 800 1060 hPa 20 C 70 C 20 75 800 1060 hPa 5 6 2 3 IEC60601 1 20...

Page 106: ...4 6 m MMAD 0 35 ml min 6 ml 23 dB 1 NGI Next Generation EN 13544 1 2007 A1 2009EN 13544 1 2007 A1 2009 0 20 40 60 80 100 0 1 1 10 100 0 20 40 60 Aerodynamic diameter um Mass fraction Cumulative mass...

Page 107: ...107 BG EN60601 1 2 AGU N7 MBPN002 EN60601 1 2 2007 AGU N7 MBPN002 AGU N7 AGU N7 CISPR 11 1 AGU N7 CISPR 11 AGU N7 61000 3 2 61000 3 3...

Page 108: ...8 AGU N7 AGU N7 60601 61000 4 2 6 8 6 8 30 61000 4 4 2 1 2 61000 4 5 1 2 1 61000 4 11 5 UT UT 95 40 UT UT 60 5 70 UT UT 30 25 5 UT UT 95 5 c 5 UT UT 95 40 UT UT 60 5 70 UT UT 30 25 5 UT UT 95 5 c AGU...

Page 109: ...109 BG 50 60 61000 4 8 3 A 3 A AGU N7 61000 4 6 61000 4 3 3 150 80 3 80 2 5 3 3 AGU N7 d 1 2 P d 1 2 P 80 800 d 2 3 P 800 2 5 P d a b...

Page 110: ...80 MHz 800 MHz 2 a FM AGU N7 AGU N7 AGU N7 150 80 M 3 AGU N7 AGU N7 AGU N7 AGU N7 150 80 d 1 2 P 80 800 d 1 2 P 800 2 5 d 2 3 P 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73 1 1 2 1 2 2 3 10 3 8 3 8 7 3 100...

Page 111: ...111 BG d m P W 1 80 MHz 800 MHz 2 2 www agu baby com...

Page 112: ...sence de l acheteur The set completeness and operation of the device are tested in the presence of the buyer Name vornamedesk ufers Nom de l acheteur Name of purchaser Unterschri des k ufers Signature...

Page 113: ...ATE GRUND CAUSE DE L APPEL REASON EMPFEHLUNGEN RECOMMANDATIONS RECOMMENDATIONS DATUM DATE DATE GRUND CAUSE DE L APPEL REASON EMPFEHLUNGEN RECOMMANDATIONS RECOMMENDATIONS Anmeldung bei Service und Wart...

Page 114: ...r Verdauungssystem gelan gen k nnen Veillez ce que les enfants ne puissent pas utiliser l appareil sans surveillance parce que certaines de ses petites parties peuvent p n trer dans les voies respi ra...

Page 115: ...AGU is the registered trademark by Montex Swiss AG Tramstrasse 16 CH 9442 Berneck Switzerland...

Reviews: