background image

LD-210С

RUS

Ингалятор

 

компрессорный

 LD

Руководство

 

по

 

эксплуатации

KAZ

ENG

UKR

Компресорний

 

інгалятор

 LD 

Інструкція

 

з

 

експлуатації

Компрессорлы

 

ингаляторы

 LD

Пайдалану

 

жөніндегі

 

басшылық

 

құжат

Compressor Nebulizer LD

Instruction Manual

Summary of Contents for LD-210C

Page 1: ...LD 210 RUS LD KAZ ENG UKR LD LD Compressor Nebulizer LD Instruction Manual...

Page 2: ...ENG 38 PARTS AND COMPONENTS 2 1 3 9 10 11 2 4 5 8 6 7 15 14 16 17 18 19 13 22 B 12 1 12 7 12 6 12 4 12 5 12 A 12 2 C 12 3 B 12 4...

Page 3: ...affle mainly the lower respiratory tract Consist of two parts cone and cylinder Expendable material 12 3 Baffle C Red LD N003 Inhalation baffle mainly the upper respiratory tract Expendable material 1...

Page 4: ...solution aerosols This Instruction Manual is designed to assist the user with safe and effective operation of the Compressor Nebulizer LD Use this Device according to the rules described in this Manu...

Page 5: ...mponents to strong shocks such as dropping on the floor This device is approved for human use only Do not disassemble or attempt to repair the device or components Use the device only for its intended...

Page 6: ...ot wrap the power cord around the compressor main unit The followings are maintenance and repair which can be taken by operator or which must be oper ated by manufacturer or distributor SERVICE AND MA...

Page 7: ...affle C For a more effective impact on the lower respiratory tract baffle A yellow color which consists of two parts Fig 2 figures indicate the order of assembly Graphics of the differential particle...

Page 8: ...total mass 1 100 60 30 10 7 5 3 0 5 10 15 particle sizes mkm to the total mass 1 100 60 30 10 7 5 3 0 5 10 15 Measured by Little Doctor International S Pte Ltd Performing the inhalation The length of...

Page 9: ...t Clean and wash the device as described in last paragraph SAFETY INFORMATION WARRANTY The following LD product is covered by warranty for the period 36 months The warranty does not apply to the consu...

Page 10: ...nutes on 40 minutes off Operation Temperature and humidity atmospheric pressure 10 C 35 C 15 80 860 1060hPa Transport and storage temperature and humidity atmospheric pressure 20 C 40 C 15 95 500 1060...

Page 11: ...g use Turn the device off Wait 40 minutes before using the device again CERTIFICATION AND STATE REGISTRATION This device manufacturing is certified according to international standard ISO 13485 Device...

Page 12: ...Little Doctor International S Pte Ltd Little Doctor International S Pte Ltd 2020 I408 2010 26 UA TR 001 WWW LITTLEDOCTOR SG...