64
NOTE:
UT is the a.c. mains voltage prior to application of the test level.
GUIDANCE AND MANUFACTURER’S
DECLARATION ELECTROMAGNETIC IMMUNITY
Mesh–nebulizer (membrane)
AGU N7
is intended for use in the electromagnetic environ-
ment specified below. The customer or the user of the mesh–nebulizer (membrane)
AGU N7
should assure that it is used in such an environment.
Immunity test
IEC 60601
test level
Compliance level
Electromagnetic
environment–guidance
Electrostatic
discharge (ESD)
IEC 61000–4–2
+ 6 kV contact
+ 8 kV air
+ 6 kV contact
+ 8 kV air
Floors should be wood,
concrete or ceramic tile.
If floors are covered with
synthetic material,
the relative humidity
should be at least 30 %
Electrical fast
transient / burst
IEC 61000–4–4
+ 2 kV for power
supply lines
+ 1 kV for
input / output lines
+ 2 kV for power
supply lines
Not applicable
Main power quality
should be that
of a typical commercial
or hospital environment
Surge
IEC 61000–4–5
+ 1 kV line(s)
to line(s)
+ 2 kV line(s)
to earth
+ 1 kV differential
mode
Not applicable
Main power quality
should be that of a typical
commercial or hospital
environment
Voltage Dips,
short interrup-
tions and volt-
age variations
on power supply
input lines
IEC 61000–4–11
<5 % UT
(>95 % dip in UT)
for 0,5 cycle
40 % UT
(60 % dip in UT)
for 5 cycles
70 % UT
(30 % dip in UT)
for 25 cycles
<5 % UT
(>95 % dip in UT)
for 5 sec.
<5 % UT
(>95 % dip in UT)
for 0,5 cycle
40 % UT
(60 % dip in UT)
for 5 cycles
70 % UT
(30 % dip in UT)
for 25 cycles
<5 % UT
(>95 % dip in UT)
for 5 sec.
Main power quality
should be that of a typical
commercial or hospital
environment. If the user
of the
AGU N7
requires
continued operation
during power mains
interruptions, it is recom-
mended that the
AGU N7
be powered from an
uninterruptible power
supply or a battery
Power frequen-
cy (50, 60 Hz)
magnetic field
IEC 61000–4–8
3 A / m
3 A / m
The
AGU N7
power
frequency magnetic
fields should be at levels
characteristic of a typi-
cal location in a typical
commercial or hospital
environment
Summary of Contents for Minimill N7
Page 68: ...68 1 69 2 70 3 71 4 72 5 73 6 75 7 78 8 80 9 82 10 83 11 84 12 89 AGU N7 RU...
Page 69: ...69 RU AGU N7 AGU N7 AGU N7 AA AGU N7 AGU N7 Active Vibrating Mesh AGU N7...
Page 70: ...70 II IEC 60601 1 BF IEC 60601 1 12 5 15 ON OFF DC 93 42 EEC 020 2011 004 2011 WEEE...
Page 71: ...71 RU M 10 C 40 C 30 85...
Page 72: ...72...
Page 73: ...73 RU 1 2 AGU MC7 AGU MC7 4 3...
Page 74: ...74...
Page 75: ...75 RU AGU N7 AA NiMH IEC 62133 1 AA NiMH...
Page 76: ...76 3 6 4 2...
Page 77: ...77 RU 5 AGU N7 4 AA 20 AA AA NiMH...
Page 78: ...78 IEC60601 1 2005 DC 5V 1A 3 5 1 35 9 5 1 2 6...
Page 79: ...79 RU 3 4 5...
Page 80: ...80 1 2 1 5...
Page 81: ...81 RU 2 5...
Page 82: ...82...
Page 89: ...89 RU d P 1 80 800 2 2 agu baby com...
Page 90: ...90 1 91 2 92 3 93 4 94 5 95 6 97 7 100 8 102 9 104 10 105 11 106 12 111 AGU N7 BG...
Page 91: ...91 BG AGU N7 AGU N7 AGU N7 AGU N7 AGU N7 Active Vibrating Mesh AGU N7...
Page 92: ...92 II IEC 60601 1 BF IEC 60601 1 12 5 15 ON OFF DC 93 42 EEC 020 2011 004 2011 WEEE...
Page 93: ...93 BG 10 40 30 85...
Page 94: ...94...
Page 95: ...95 BG 1 2 AGU MC7 AGU MC7 4 3...
Page 96: ...96...
Page 97: ...97 BG AGU N7 AA NiMH IEC 62133 1 NiMH...
Page 98: ...98 3 6 4 2...
Page 99: ...99 BG 5 AGU N7 4 20 NiMH...
Page 100: ...100 IEC 60601 1 2005 AC 5V 1A 3 5 mm 1 35 9 5 1 2 6...
Page 101: ...101 BG 3 4 5...
Page 102: ...102 1 2 1 5...
Page 103: ...103 BG 2 75 1 5...
Page 104: ...104...
Page 111: ...111 BG d m P W 1 80 MHz 800 MHz 2 2 www agu baby com...
Page 115: ...AGU is the registered trademark by Montex Swiss AG Tramstrasse 16 CH 9442 Berneck Switzerland...