Specification
3
Agilent U9391C/F Comb Generator Operating and Service Manual
27
Environmental Specifications
Agilent U9391C/F are designed to fully comply with Agilent
Technologies’s product operating environment specifications.
The following are the summarized environmental specifications
for these product.
Table 3-4
U9391C/F Environmental Specifications
Temperature
•
Operating
•
Storage
•
Error corrected range
•
Cycling
0
°
C to +55
°
C
-40
°
C to +70
°
C
23
°
C to +/- 3
°
C
-65
°
C to +85
°
C, 10 cycles at 20
°
C per minute. 20 minutes dwell time per
MIL-STD-883F, Method 1010.8, Condition C (modified)
Relative Humidity
•
Operation
•
Storage
50% to 95% RH at 40
°
C, 24 hours cycling, repeated 5 times
90% RH at 65
°
C, one 24 hour cycle
Shock
•
End-use handling shock
•
Transportation shock
Half-sine wave form, 2-3ms duration, 60 in/s (1.6ms) delta-V
Trapezoidal wave form, 18-22ms duration, 337 in/s (8.56ms) delta-V
Vibration
•
Operating
•
Survival
Random: 5-500Hz, 0.21grms, 10min/axis
Random: 5-500Hz, 2.09grms, 10min/axis
Swept Sine: 5-500Hz, 0.5grms, 10min/axis, 4 resonance search, 10min dwell
Altitude
•
Operating
•
Storage
< 4,572 meters (15,000 feet)
< 15,000 meters (50,000 feet)
ESD immunity:
•
Direct discharge
*
•
Air discharge
* To outer conductor
8.0 kV per IEC 61000-4-2
15 kV per IEC 61000-4-2
Summary of Contents for U9391C
Page 1: ...Agilent Technologies Agilent U9391C F Comb Generator Operating and Service Manual ...
Page 6: ...6 Agilent U9391C F Comb Generator Operating and Service Manual ...
Page 12: ...12 Agilent U9391C F Comb Generator Operating and Service Manual 1 Introduction ...
Page 30: ...30 Agilent U9391C F Comb Generator Operating and Service Manual 3 Specification ...