Agilent U9391C/F Comb Generator Operating and Service Manual
37
12
Connect the
Output
of phase reference unit (unit that has
been selected as phase calibration) to
Port 3
of network
analyzer once calibration is completed.
13
Click on the
Measurement Display
tab and select as
following to perform comb amplitude and comb phase
measurement.
a
Format
: select
Log Mag
(to measure comb amplitude) or
Phase
(to measure comb phase)
b
Domain
: select
Frequency
c
Measure
: select
Single
or
Continuous
Figure 4-5
NVNA Calibration Wizard