Agilent U9391C/F Comb Generator Operating and Service Manual
2
Installation
The phase calibration procedure is carried out by following the
on- screen instructions on the PNA- X.
After completing the calibration, the U9391C/F unit used for
phase calibration can be disconnected. The other unit that is
used as the phase reference will remain connected while the
measurement is carried out.
For more detailed information, please refer to the Help file that
is embedded in the NVNA application.
C A U T I O N
When connecting the U9391C/F directly to the instrument test ports,
the modules will require mechanical support so as not to overstress
the test port connectors.