Performance Tests
7
Agilent U1602A/U1604A User’s and Service Guide
169
Table 7-5
DC Volts Verification Test
Figure 7-6
Test connections from calibrator to instrument
Calibrator Voltage
Function
Range
Min Limit
Max Limit
600 mV
DC Volts
600 mV
597.7 mV
602.3 mV
6 V
6 V
5.977 V
6.023 V
30 V
60 V
29.86 V
30.14 V
60 V
60 V
59.77 V
60.23 V
600 V
600 V
597.7 V
602.3 V