7
Performance Tests
172
Agilent U1602A/U1604A User’s and Service Guide
Capacitance Verification Test
1
Put the calibrator in Standby (STBY) mode.
2
Select Meter mode of the instrument, press F2 softkey for Ohmmeter.
3
Press F1 to select capacitance measurement.
4
Connect the calibrator to the instrument’s input and ground terminal as
shown in Figure 7- 2 using a banana plug.
5
Select each function and range in the order shown in Table 7- 8. Apply
the values from the calibrator shown in table 7- 8 below.
6
Compare the measurement results to the corresponding test limits
shown in table 7- 8.
Table 7-8
Capacitance Verification Test
Calibrator Settings
Function
Range
Min Limit
Max Limit
60 nF
Capacitance
60 nF
58.68 nF
61.32 nF
600 nF
600 nF
586.8 nF
613.2 nF
6
m
F
6
m
F
5.868
m
F
6.132
m
F
36
m
F
60
m
F
35.16
m
F
36.84
m
F
60
m
F
60
m
F
58.68
m
F
61.32
m
F
300
m
F
300
m
F
293.4
m
F
306.6
m
F