Performance Tests
7
Agilent U1602A/U1604A User’s and Service Guide
173
Diode Verification Test
1
Put the calibrator in Standby (STBY) mode.
2
Select Meter mode of the instrument, then press F2 softkey for
Ohmmeter.
3
Press F1 to select diode test.
4
Connect the calibrator to the instrument’s input and ground terminal as
shown in Figure 7- 2 using a banana plug.
5
Select each function and range in the order shown in Table 7- 9. Apply
the values from the calibrator shown in Table 7- 9 below.
6
Compare the measurement results to the corresponding test limits
shown in table 7- 9.
Table 7-9
Diode Verification Test
Calibrator Voltage
Function
Min Limit
Max Limit
0.9 V
Diode
0.8812 V
0.9188 V