Technical Information
Window Meanings
17
Turbo-V 1001 Rack Controller User Manual / 87-901-033-01
239/288
N.
Read/
Write
Data
Type
Description
Admitted Values
106 R/W
L
Cooling agent
0 = AIR
1 = WATER
(default = 0)
107 R/W
L
Active Stop
(can be written only in Stop
condition)
0 = NO
1 = YES
(default = 0)
108 R/W
N
Baud rate
600 = 0
1200 = 1
2400 = 2
4800 = 3
9600 = 4
19200 = 5
38400 = 6
(default = 4)
110 R/W
L
Interlock mode (if configured
as Impluse the Interlock signal
acts only at the start)
Impulse = 0
Continuous = 1
(default = 1)
111 R/W
L
Analog output link: the output
voltage can be related to
different quantities.
0 = frequency
1 = power
2 = bearing temp
(default = 0)
117 R/W
N
Low speed setting [Hz]
350 to WIN 121
(default = 750)
120 R/W
N
High speed setting [Hz]
350 to WIN 121
(default = 825)
121 R/W
N
Maximum rotational frequency
in Hz (can be written only in
Stop condition)
350 to 825
(default = 825)
122 R/W
L
Set Vent Valve status
1 = open
0 = close
(default = 0)
125 R/W
L
Valves operating mode
0 = On Command
1 = Auto Pump Spedd
2 = Autotime
(default = 2)