17 Technical Information
Controller Specifications
212/288
Turbo-V 1001 Rack Controller User Manual / 87-901-033-01
Controller Specifications
Tab. 2
Input:
Voltage
Frequency
Power
100 ÷ 240 Vac (±10 %), 1-phase
50 to 60 Hz
800 VA maximum
Output for pump:
Power during pump ramp-up
Power normal (pump water cooled)
Power normal (pump air cooled)
450 W maximum
450 W maximum
300 W maximum
Operating temperature
5 °C to +45 °C
Storage temperature
-20 °C to +70 °C
Fuse:
Mains
2 x T10 AH (slow blow) 250 V 5x20 fuses
Serial communication (T-Plus kit)
RS232 cable with a 9-pin D type male connector and a 9-
pin D type female connector, and T-Plus software
(optional)
Power cable
With European or NEMA plug 3 meters long (optional)
Pump cable
5 meters long fixed cable
In compliance with norms
EN 61010-1 (2001)
EN 61326-1 (2006)
Emission level : Class B for Residential Application
Immunity level : for Industrial Application
Max altitude
2000 m
Protection category
IP 20
Internal use only
Weight (both models)
6.3 kg (13.9 lbs)
Installation category
II
Pollution degree
2