17 Technical Information
Window Meanings
240/288
Turbo-V 1001 Rack Controller User Manual / 87-901-033-01
N.
Read/
Write
Data
Type
Description
Admitted Values
126 R/W
N
Vent valve opening delay
expressed in 0.2 sec
0 to 65535
Corresponding to 0 to
13107 sec
(Default = 15)
143 R/W
N
External Fan Configuration
0 = Always ON
1 = Automatic
2 = Serial
(default = 2)
144 R/W
L
External Fan activation
0 = OFF
1 = ON
(default = 0)
145 R/W
L
Purge Valve open/close (N.C.) 0 = close
1 = open
(Default =1)
147 R/W
N
Vent Valve opening (Time
expressed in 0.2 s)
0 to 65535 bit
0 = infinite
min 0.2 s,
Max 13107 s
(default = 0)
155 R
N
Power Limit Applied expressed
in W
157 R/W
L
Gas Load Type
0 = Ar
1 = N
2
(default = 0)