16
Site Preparation
Temperature
Requirements
The Agilent E22xxA Test System is designed to operate in the range from
5° C to 40° C (41° F to 104° F).
Cooling Requirements
As an option to the system, the system rack can be equipped with three
extractor fans which draw air through the rack and exhaust it through the top
rear of the rack. With the rack side panels and top installed, the fans must be
operational to prevent excessive heat rise inside the rack.
Caution
As a convenience, the switch marked ‘Fan Switch Only’ can be
switched off during maintenance when the rack side panels and
top are removed. To prevent equipment damage from
overheating, the fans should be turned on at all other times
when the rack side panels and top are in place.
Humidity
Requirements
The test system is designed to operate in the range from 5 percent to 80
percent relative humidity (non-condensing).
If the system is subjected to condensation, as when moved from a cold
loading dock into a warm (and damper) environment, allow at least 24 hours
for the system to dissipate the condensation before powering up.
Pollution Degree
Degree 2 (IEC 664) (Indoor use only)
Maximum Altitude
2000 meters