background image

Site Preparation  

13

Chapter 1

*The approximate weight of a fully-loaded system is 112 kilograms (248 pounds) excluding the 

pallet and shipping crate. The pallet and shipping crate, together, weigh approximately 90 
kilograms (200 pounds).

Moving Access

Requirements

The system is shipped from the factory on one pallet in a plywood shipping 
crate. The pallet may be too large to move directly to your system's 
destination.

Ramp Requirements

The rack has casters and will negotiate ramps with inclines up to 8 percent 
before the leveling feet drag at the ramp transition.

Hallway and Door Width

Requirements

Make sure that all doors, elevators and passageways leading to the system’s 
final location are large enough to allow passage of the crated system. The 
crated system requires the use of a floor jack or fork lift to engage the pallet 
and lift the crated system. The dimensions of the system crate including 
pallet are 41 inches W x 42.5 inches L x 53.25 inches H. 

If the system cannot be

moved in the crate

Remove the system from the pallet in the receiving area and push it on the 
rack casters to the final destination if obstacles or lack of space restrict 
moving the crated system.

Note

The rack casters can only transition very small floor level changes or gaps. 

An uncrating document (Agilent part number E2230-90025) is attached to 
the outside of the crate and describes how to uncrate the system.

Table 1-1. E22xxA Functional Test System Rack Specifications

Size

Dimension

Height

Width

Depth

EIA

0.75m

Exterior

740mm (29.1in)

600mm (23.6in)

Outside of rack: 750mm (29.5in)
Including Express Connect, PC and 
Power Supply overhangs: 850mm 
(33.46in)

15 RU

Summary of Contents for TS-5410

Page 1: ...Agilent Technologies Agilent TS 5410 Compact Functional Test System Site Preparation and Installation Guide...

Page 2: ...document are furnished under a license and may be used or copied only in accordance with the terms of such license Restricted Rights Legend If software is for use in the performance of a U S Governmen...

Page 3: ...e protection provided by the equipment may be impaired Cleaning Instructions Clean the system cabinet using a soft cloth dampened in water WARNING DO NOT REMOVE ANY SYSTEM COVER Operating personnel mu...

Page 4: ...y equipment in a locked cabinet is required to gain access to them Additionally the operator cannot have access to a conductive surface connected to any cable conductor High Low or Guard ASSURE THE EQ...

Page 5: ...letely disconnected from ac mains when this switch is off To completely disconnect the unit from ac mains either disconnect the power cord or have a qualified electrician install an external switch Re...

Page 6: ...vice on the Web http www agilent com Click the link to Test Measurement Select your country from the drop down menus The Web page that appears next has contact information specific for your country Ag...

Page 7: ...7 Doc Placeholder Page...

Page 8: ...8...

Page 9: ...ticulate Contaminants 15 Viscid Contaminants 15 Cooling Requirements 16 The Site Plan Drawing 17 AC Power Requirements 18 Important Information 18 Alternate Disconnect Recommendation 19 Optional Emerg...

Page 10: ...10 Contents...

Page 11: ...4 Environmental Requirements page 15 The Site Plan Drawing page 17 AC Power Requirements page 18 Alternate Disconnect Recommendation page 19 System Intended Use Agilent E22xxA systems are designed to...

Page 12: ...t Load Box Other Mass Connect 5 10 DCV 0 74 m 0 6 m 0 75 m 0 06 m DC Fan Assy FRONT REAR PC PC Interconnect AC Power Mass Connect AC Power 34401A DMM 53131A Counter 33220A Arb Other 3499A B Switch Con...

Page 13: ...ge of the crated system The crated system requires the use of a floor jack or fork lift to engage the pallet and lift the crated system The dimensions of the system crate including pallet are 41 inche...

Page 14: ...rack is secured to the pallet of the shipping crate and wrapped with a plastic wrap Do not move the crate or the test rack and pallet to a static sensitive area until you have removed the plastic wra...

Page 15: ...al are complex and costly If the source of corrosive contaminants cannot be eliminated the system should be installed in an enclosed environment with a fresh air supply at positive pressure Particulat...

Page 16: ...ce the switch marked Fan Switch Only can be switched off during maintenance when the rack side panels and top are removed To prevent equipment damage from overheating the fans should be turned on at a...

Page 17: ...If the operator will be positioned in front of the system allow 1 meter 3 3 feet of space in front of the system You should allow 1 meter 3 3 feet of space behind the system for service The 0 75 mete...

Page 18: ...endations The system is to be supplied from a dedicated 15 16 Amp single phase service at the voltage for which the system is configured The system is delivered with a mating power connector for the s...

Page 19: ...ndation Emergency power off EMO capability is not provided with the Filtered Power Strip An external mains disconnect switch or breaker should be installed adjacent to the equipment for easy access Us...

Page 20: ...20 Site Preparation Chapter 1...

Page 21: ...nd how to backup the system software Chapter contents are Grounding and Connecting Power to the System page 22 PC and Peripheral Connections page 25 Fixture and UUT Connections page 27 Getting Started...

Page 22: ...position Lower the test rack leveling feet to stabilize the rack and prevent it from rolling 2 Make a permanent connection from the system rack to protective earth ground This connection will serve a...

Page 23: ...minated power cord Wiring information for this cord is shown in Figure 2 3 Figure 2 1 Safety Ground Connections SystemRack T oothed Washers WireLug Earth ground terminal Protective Earth connection at...

Page 24: ...g type needed and if necessary the power cord needed Make the appropriate connections between the cord and wall outlet connector and the Power Connector 5 Set the power switch on the top front of the...

Page 25: ...ral Connections For your convenience PC connections have been brought out to the rear of the system on the PC and GPIB Connector Panel Figure 2 5 1 Connect the monitor keyboard mouse LAN and any USB o...

Page 26: ...setting the front panel on off switch to ON Note DO NOT secure the AC power cord to the cable strain relief bracket such that it would prevent the plug from being removed from the power strip in an e...

Page 27: ...d mark can be seen on the breaker switch Press the breaker button to reset Fixture and UUT Connections The Test System Interface on the front of the system allows you to make connections between the t...

Page 28: ...luded with your system This guide shows you how to make connections to the system s Test System Interface This manual is also available from Test Exec SL by clicking Tools TS 5400 Online Manuals This...

Page 29: ...power to the system 22 H Hallway and Door Width Requirements 13 Hole for routing wires 25 Humidity Requirements 16 I IEC320 C19 connector 23 If the system cannot be moved in the crate 13 K Keyboard 25...

Page 30: ...k 12 Site plan 17 Static precautions 14 System Power Options 19 System Weight 13 T Temperature Requirements 16 Top view rack 12 Typical Power Plug Connections 23 Typical test system layout overhead vi...

Reviews: