18
Site Preparation
AC Power Requirements
Agilent E22xxA Automotive Functional Test Systems are equipped with a
Filtered Power Strip that distributes AC Mains power to the various system
equipment. Figure 2-3 shows the location of the Filtered Power Strip. The
standard system is configured for 230-240 VAC, single-phase, 50/60 Hz.
The system can be configured for other single-phase voltages by contacting
Agilent service personnel. Some instrumentation within the system must be
reconfigured. The possible voltage ranges and currents are as follows:
•
100 VAC, 8 A,
•
110 - 120 VAC, 8 A,
•
200 - 220 VAC, 4 A,
•
230 - 240 VAC, 4 A.
System Power
Recommendations
The system is to be supplied from a dedicated 15/16 Amp single-phase
service at the voltage for which the system is configured.The system is
delivered with a mating power connector for the system appliance inlet.
A customer supplied power cord must meet the requirements of the country
where the system is installed.
A supplemental Protective Earth ground wire is required for leakage current.
Caution
Verify the AC source and that service conductors are sized
correctly before connecting the system.
Important Information:
•
Conventional 50/60-hertz current probes will not make accurate input
current measurements of the system’s power supplies due to the
harmonics on the currents. A current measuring instrument needs a
bandwidth that is above 10 kHz. Improper instruments may yield
results that are 50 percent less than actual.