Command Reference
62
Agilent Sample Introduction Pump System SIPS 10/20 User’s Guide
Dilute Sample By
Select this checkbox and enter an initial dilution factor to be used for
all samples.
Overrange dilution can be performed online using SIPS. If it is
expected that samples are over the calibration range, SIPS
automatically uses this dilution factor to dilute all samples before
measurement.
Dual Pump Recalibration Group
This group is enabled only when the selection for the
Right Pump
group has been set to
Bulk Standard
. It allows you to periodically
recalibrate the pumps during a long autorun, to correct for any
possible drift which may have occurred, particularly if there is any
significant difference in the ages of the sample pump tubing and the
standard pump tubing.
NOTE
If the periodic pump recalibration fails, the error message “Periodic pump
recalibration failed" appears and the calibration error action selected on the
Options
page of the
Sequence Parameters
window will be executed.
With each Recalibration
— Select this checkbox if you want a pump
recalibration to occur at the start of each recalibration.
With each Reslope
— Select this checkbox if you want a pump
recalibration to occur at the start of each reslope.
Standards Page
The following controls are relevant to SIPS: