Specifications
Agilent Sample Introduction Pump System SIPS 10/20 User’s Guide
21
3.
Specifications
Environmental 21
Power 22
External Connections
22
Weights and Dimensions
23
Performance 23
Environmental
Your accessory is designed for indoor use only. In addition, this
accessory is suitable for the following categories:
Installation category II
Pollution degree 2
Equipment class I
Condition
Altitude
Temp t (°C) Humidity (%RH)
non-condensing
Non-operating (transport)
0–2133 m (0-7000 ft)
5–45
20–80
Operating within
performance specifications
0–853 m (0–2800 ft)
853–2133 m (2800–7000 ft)
10–35
10–25
8–80
NOTE
For optimum analytical performance, it is recommended that the ambient
temperature of the work area be between 20 and 25 °C (68 and 77 °F) and be
held constant to within ±2 °C throughout the entire working day.