Installation
Agilent Sample Introduction Pump System SIPS 10/20 User’s Guide
27
NOTE
The serial number of the pump unit is on the underside of the control module.
The serial number of the electronic control module can be found on its side,
next to the fuses.
Please refer to the packing list that is shipped with the SIPS to
confirm that you have received all of the components shipped.
Power
WARNING
Electrical Shock – Fire Hazard
Application of the wrong supply voltage, connection of the accessory to
an incorrectly wired supply outlet, or lack of proper electrical
grounding can create a fire or shock hazard that can cause death,
serious injury or serious damage to equipment. Do not connect the SIPS
to the mains power until you have checked the power requirements as
specified on Page 22.
Checking the Fuses
The stand alone electronic control module contains two 5 × 20 mm
IEC fuses, located on the side of the module along with the various
connection ports.
NOTE
The internal power supply module (280/240 series AA) fuse can only be
replaced by an Agilent field service engineer.
Fuses have a code marked on the cap (e.g., T2AH250V). This refers to
the fuse characteristic (T - time lag, F - fast acting), the current rating
(x amperes), the breaking capacity (H - heavy, L - low), and the
voltage rating (y volts).