Installation
Agilent Sample Introduction Pump System SIPS 10/20 User’s Guide
31
To connect the SIPS unit to the electronic control module:
1
Connect one end of this cable to the appropriate port on the
electronic control module.
2
Secure the cable in place by sliding the metal bracket on the
connector to the right, over the studs located on either side of the
connector port.
3
Connect the other end of the cable to the port located underneath
the pump unit on the right-hand side, sliding the bracket on the
connector to one side to secure it in place. The cable may be
routed around the bottom edge of the instrument, or underneath
it.
CAUTION
The SIPS is not suitable for use with organic solvents because the tubing is not
resistant to these solvents.
NOTE
If you are using an Agilent 280FS AA or 140/240 AA
4
, the power for the
electronic control module is supplied from the spectrometer. The following
steps in this procedure are not applicable.
4
The SIPS is shipped with three power cables. Select the
appropriate one for your location.
5
Ensure that the power switch on the electronic control module is
turned off (to ‘0').
6
Plug one end of the power cable into the electronic control
module and the other end into the mains power outlet.
7
You may now turn the SIPS on by pressing the power switch to
‘I'. The green indicator lamp should light.