Installation
40
Agilent Sample Introduction Pump System SIPS 10/20 User’s Guide
6
Clean the plastic pump band by wiping it with a tissue moistened
with alcohol.
7
Using an alcohol-moistened cloth, wipe the pump rollers. Ensure
they all free of corrosion and dirt, and rotate freely.
8
Clean the nebulizer and Pt/Ir capillary using the standard
procedure shown in the SpectrAA Online Help - Cleaning a
Blocked Nebulizer.
9
Assemble the SIPS system and install the pump tube in place.
10
Now condition the pump tube as described on Page 38.