144
Storing, Logging and Printing
Logging Results
•
External Parallel Port printer (option 601)
•
Internal Disk Drive
HOW TO:
1
Choose LOGGING [ON] - enables the logging of results and alarms.
2
Choose LOGGING PERIOD - determines how regularly the results and alarms
are logged.
USER PROGRAM provides a choice of 10 minutes to 99 hours.
3
Choose RESULT LOGGED - allows you to log all results to or choose only those
results you require.
4
Choose WHEN - allows you to choose to only log when the error count for the
logging period is greater than 0. If the error count is 0 then the message NO BIT
ERRORS is displayed.
5
Choose CONTENT - allows you a choice of error results to be logged.
Error Results, Analysis or Error and Analysis (ER & ANAL)
and
Period, Cumulative or Period and Cumulative (PER & CUMUL).
6
If LOG ERROR SECONDS [ON] is chosen a timed and dated message is logged
each time an error second occurs (excessive occurrences of error seconds during
the logging period will result in heavy use of printer paper).
7
Choose the logging DEVICE.
If RS232 is chosen, see "Logging Results to RS-232-C Printer " page 151.
If GPIB is chosen, see "Logging Results to GPIB Printer " page 149.
If PARALLEL is chosen, see "Logging Results to Parallel (Centronics) Printer
" page 148.
SONET_Usr.bk Page 144 Wednesday, December 13, 2000 11:11 AM
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com