OUTPut Commands
OUTPut Commands
Output commands control the output, power-on, and protection clear functions.
OUTPut [:STATe] 0|OFF|1|ON
OUTPut[STATe]?
Enables or disables the output. The state of a disabled output is a condition of zero output voltage and zero source cur-
rent.
Parameter
Typical Return
0|OFF|1|ON , *RST OFF
0 or 1
Turns the output off: OUTP OFF
l
When output is enabled, the front-panel status indicator changes from
OFF
to indicate the operating status of the
instrument (
CV
,
CC
, etc.).
OUTPut:PON:STATe RST|RCL0
OUTPut:PON:STATe?
Sets the output power-on state. This determines whether the power-on state is set to the *RST state (RST) or the
state stored in memory location 0 (RCL0).Instrument states can be stored using the *SAV command. This parameter
is saved in non-volatile memory.
Parameter
Typical Return
RST|RCL0
RST or RCL0
Sets the power-on state to the *RST state: OUTP:PON:STAT RST
OUTPut:PROTection:CLEar
Resets the latched protection. This clears the latched protection status that disables the output when a protection con-
dition occurs (see
Programming Output Protection
).
Parameter
Typical Return
(none)
(none)
Clears the latched protection status: OUTP:PROT:CLE
l
All conditions that generate the fault must be removed before the latched status can be cleared. The output is
restored to the state it was in before the fault condition occurred.
82
Agilent N8900 Series Operating and Service Guide