Installation Note N5230-90022 5
Installation Procedure for the Upgrade
The network analyzer must be in proper working condition prior to installing this option. Any
necessary repairs must be made before proceeding with this installation.
WARNING
This installation requires the removal of the analyzer’s protective outer
covers. The analyzer must be powered down and disconnected from the
mains supply before performing this procedure.
Electrostatic Discharge Protection
Protection against electrostatic discharge (ESD) is essential while removing or connecting
cables or assemblies within the network analyzer.
Static electricity can build up on your body and can easily damage sensitive internal circuit
elements when discharged. Static discharges too small to be felt can cause permanent
damage. To prevent damage to the instrument:
•
always
have a grounded, conductive table mat in front of your test equipment.
•
always
wear a grounded wrist strap, connected to a grounded conductive table mat, having
a 1 M
Ω
resistor in series with it, when handling components and assemblies or when
making connections.
•
always
wear a heel strap when working in an area with a conductive floor. If you are
uncertain about the conductivity of your floor, wear a heel strap.
•
always
ground yourself before you clean, inspect, or make a connection to a static-sensitive
device or test port. You can, for example, grasp the grounded outer shell of the test port or
cable connector briefly.
shows a typical ESD protection setup using a grounded mat and wrist strap. Refer to
“ESD Equipment and Supplies Required for the Installation” on page 6
Figure 1
ESD Protection Setup
Summary of Contents for N5230AU-974
Page 13: ...Installation Note N5230 90022 13 Figure 8 20 GHz Top to Bottom Cables Removal Top...
Page 15: ...Installation Note N5230 90022 15 Figure 9 20 GHz Components Removal...
Page 17: ...Installation Note N5230 90022 17 Figure 10 50 GHz Components Installation...
Page 19: ...Installation Note N5230 90022 19 Figure 12 50 GHz Top to Bottom Cables Installation Top...
Page 23: ...Installation Note N5230 90022 23 Figure 15 50 GHz Bottom Cables Installation...