![Agilent Technologies N5230AU-974 Installation Note Download Page 10](http://html1.mh-extra.com/html/agilent-technologies/n5230au-974/n5230au-974_installation-note_2868706010.webp)
10
Installation Note N5230-90022
Step 3. Remove the 20 GHz Bottom Cables
Refer to
for this procedure.
1. Place the analyzer bottom-side up on a flat surface.
2. Remove the following cables in the order listed:
• W40 A17 MASS 26.5 to A25 port 1 step attenuator
• W42 A25 port 1 step attenuator to PORT 1 SOURCE OUT
• W44 PORT 1 CPLR THRU to A21 test port 1 coupler
• W46 A17 MASS 26.5 to REFERENCE 1 SOURCE OUT
• W51 REFERENCE 1 RCVR R1 IN to A20 mixer brick (R1)
• W48 A21 test port 1 coupler to PORT 1 CPLR ARM
• W50 PORT 1 RCVR A IN to A20 mixer brick (A)
• W41 A17 MASS 26.5 to A26 port 2 step attenuator
• W43 A26 port 2 step attenuator to PORT 2 SOURCE OUT
• W45 PORT 2 CPLR THRU to A22 test port 2 coupler
• W47 A17 MASS 26.5 to REFERENCE 2 SOURCE OUT
• W52 REFERENCE 2 RCVR R2 IN to A20 mixer brick (R2)
• W49 A22 test port 2 coupler to PORT 2 CPLR ARM
• W53 PORT 2 RCVR B IN to A20 mixer brick (B)
Figure 5
20 GHz Bottom Cables Removal