Installation Note N5230-90022 7
Step 1. Remove the Outer and Inner Covers
Remove the Outer Cover
CAUTION
This procedure is best performed with the analyzer resting on its front
handles in the vertical position.
Do not place the analyzer on its front panel
without the handles
. This will damage the front panel assemblies.
Refer to
for this procedure.
1. Disconnect the power cord (if it has not already been disconnected).
2. With a T-20 TORX driver, remove the strap handles (item
①
) by loosening the screw
(item
②
) on each end of each handle.
3. With a T-20 TORX driver, remove the four rear panel feet (item
③
) by removing the center
screws (item
④
).
4. Slide the four bottom feet (item
⑤
) off the cover.
5. Slide the cover off of the frame.
Figure 2
Outer Cover Removal