Installation Note N5230-90022
Step 14. Enable Option F40
Procedure Requirements
• The analyzer must be powered up and operating to perform this procedure.
• The Network Analyzer program must be running.
• A mouse must be connected to the analyzer for this procedure.
Option Enable Procedure
1. On the analyzer’s
System
menu, point to
Service
, and then click
Option Enable
.
2. In the
Select Desired Option
list, click
F40 - 40 GHz Frequency Rang
e.
3. Click
Enable
.
4. Click
Yes
in answer to the displayed question in the
Restart Analyzer?
box.
5. When the installation is complete, click
Exit
.
Option Verification Procedure
Once the analyzer has restarted and the Network Analyzer program is again running:
1. On the analyzer’s Help menu, click About Network Analyzer.
2. Verify that “F40” is listed after “Options:” in the display. Click OK.
NOTE
If Option F40 has not been enabled, perform the
again. If the option is still not enabled, contact Agilent
Technologies. Refer to
Summary of Contents for N5230AU-974
Page 13: ...Installation Note N5230 90022 13 Figure 8 20 GHz Top to Bottom Cables Removal Top...
Page 15: ...Installation Note N5230 90022 15 Figure 9 20 GHz Components Removal...
Page 17: ...Installation Note N5230 90022 17 Figure 10 50 GHz Components Installation...
Page 19: ...Installation Note N5230 90022 19 Figure 12 50 GHz Top to Bottom Cables Installation Top...
Page 23: ...Installation Note N5230 90022 23 Figure 15 50 GHz Bottom Cables Installation...