8
Installation
User Guide
91
Install the LTM column modules
After installing the fan assembly, next install the selected column module. For instructions,
see:
1
To Install an LTM Column Module onto a Transfer Line Module
2
To Connect the Column to the CFT Unions
3
To Install a Column Module Assembly in the LTM Oven Door
.
Install the In and Out segments for each column
module
After installing the LTM column module, install the In and Out column segments between the
LTM column module and the GC inlet and detector. See
“To Install an In or Out Segment to
.
Check the cable routing
Check the cable routing at the bottom left corner of the LTM oven door. Make sure that some
slack is present so that the GC door can be opened at east halfway. If the cables are tight
against the GC, the LTM oven door will not open properly.
If needed, install two plastic cable clamps to the metal left side panel. One installs near the
LTM oven door, at the bottom front, and the other installs at the bottom, near the middle of the
panel.
Install the GC covers
Reinstall the GC covers.
Configure the column modules
1
Plug in the LTM system power cord(s) and the GC power cord.
2
Turn on the GC.
3
The GC will prompt you to configure the new LTM columns. Follow the prompts.
Configure each LTM column module in the GC as a composite column.