3
Column Modules
36
User Guide
•
The free column leads should extend past the two posts where the column leads exit
the module assembly. See
.
Figure 24.
Correct installation of column into transfer line
9
Secure the column module to the transfer line module using the two screws (removed in
10
Route the transfer line wires inside the screw mounts for the column module cover. See
.
Figure 25.
Routing the transfer line heater wires
Column
(wrapped in foil)
Transfer line
Poor installation.
Column exposed.
Cold spots.
Correct installation.
Column not exposed.
No cold spots.