3
Column Modules
User Guide
39
3
Trim the column.
For a new installation
, cut the column to approximately 60 mm from the face of the
transfer line assembly.
If reinstalling a column module
, inspect the column end with a magnifier. If cracked or
damaged, trim a very small length, a few millimeters, from the end.
4
Clean the column end with an alcohol wipe.
5
Slide one internal nut over the inlet column end.
Figure 27.
Internal nut and UltiMetal ferrule on column
6
Slide the appropriate UltiMetal ferrule over the inlet column end. See
7
If not already installed, slide one union hanger assembly onto the inlet transfer line arm as
shown. Be sure the hanger is at maximum extension on the arm. Use a T20 Torx driver to
tighten the screw until the hanger does not slide freely. Do not tighten completely.