![Agilent Technologies LTM II Series User Manual Download Page 40](http://html.mh-extra.com/html/agilent-technologies/ltm-ii-series/ltm-ii-series_user-manual_2867438040.webp)
3
Column Modules
40
User Guide
8
If a union hanger is already installed, slightly loosen the mounting screw so that the hanger
can slide when needed.
9
Slide the column end through a CFT union and place the CFT union into the hanger as
shown.
•
For new installations, again check that the hanger is at maximum extension without
extending past the end of the arm as shown below.
•
For existing installations, slide the hanger to reach the open column end.
CAUTION
Do not overtighten the internal nut into the fitting! A properly swaged and tightened CFT
connection will remain leak free for many connections.
End of arm
Maximum
extension
Hanger body
does not
extend past end
of arm
Lightly tighten