Column Modules
38
User Guide
To Connect the Column to the CFT Unions
Prepare the column ends and install them into the CFT unions as described below. Agilent
recommends using the CFT unions. When properly installed, the CFT unions provide reusable
leak-free connections.
You can connect the unions to the column ends before or after installing the LTM column
module into the LTM oven door. Often it is more convenient to make these connections on a
lab bench.
NOTE
Agilent strongly recommends reading this entire section before attempting to install the
union. If not familiar with Agilent's CFT fittings and unions, first practice CFT ferrule
installation on a scrap non-LTM column or column segment.
1
Gather the following:
•
Column cutter, wafer (5181-8836, 4/pk)
•
Magnifying loupe, 20X (430-1020)
•
Gloves, heat-resistant (for handling hot parts)
•
Gloves, lint free (to prevent contamination of the column, ferrules, and so forth with skin
oil and dirt)
•
UltiMetal ferrule, appropriate for the column size (see
•
Two 1/4-inch open-end wrenches. Alternate: Substitute one 6-mm wrench for a 1/4-inch
wrench.
•
T20 Torx driver
•
Short metric ruler
•
One 7/16-inch open end wrench
CAUTION
Wear clean, lint-free gloves to prevent contamination of the parts.
2
If not already done, install the column module onto the transfer line module. See