![Agilent Technologies G1888 User Information Download Page 54](http://html1.mh-extra.com/html/agilent-technologies/g1888/g1888_user-information_2867733054.webp)
54
User Information
The Keypad
Event Times
Description
Select
Event Times
from the Active Method menu to specify
the duration of various method parameters.
GC Cycle Time
The time required for the entire GC run to be
completed. It is the sum of the GC run time and GC
equilibration time (read from the GC front panel) and an
estimate of the cool–down time required (if any). The GC cycle
time must be long enough for the GC to become “ready”. This
time allows the Headspace Sampler to introduce samples into
the oven efficiently and avoid injecting a sample before the GC
is ready.
SET.
GC Cycle Time (min)
Vial Eq. Time (min)
New Setpoint:
↓
ADV. FUNCTION
27.0
Pressuriz Time (min)
25.0
15.0
0.20
SET.
Loop Fill Time (min)
Loop Eq. Time (min)
New Setpoint:
↑
ADV. FUNCTION
0.08
Inject Time (min)
0.15
0.05
0.30
medium_standard.book Page 54 Tuesday, February 17, 2004 10:14 AM