Maintaining the Instrument
User Information
135
4
Exchange the sample loop.
a
Slide the loop off the mandrel.
b
Slide the new sample loop onto the
mandrel.
•
When installing a deactivated
sample loop, take care not to bend
the loop when installing it or you
will break the coating and
compromise inertness.
5
Reverse the steps given above to
reassemble the Headspace
Sampler.
•
Perform a leak test on fittings
before replacing the insulation and
manifold cover.
Table 12
Changing the sample loop (continued)
Step
Action
Notes
Port 1
Port 4