Entering a Method
Setting Vial Parameters
Use the VIAL PARAM. screen to select sample tray locations
used in the method and enable or disable vial shaking. The
following setpoints are included:
•
First vial
•
Last vial
•
Shake
Navigate to the VIAL PARAM. screen
Press the
Active Method
key. Select
Vial Param.
from the
Active Method menu. The sampler displays the following:
For more information on vial parameters, see
on
Setting the First Vial
First Vial specifies the location of the first vial to be analyzed by
this method.
Use the cursor keys to highlight
First Vial
. Key in the
desired location, followed by
Enter
.
The allowed range for this setting is 1-70.
READY
First Vial
1
Last Vial
22
SET
VIAL PARAM.
Shake
Off