3-12
Service Guide E8364-90001
Tests and Adjustments
PNA Series Microwave Network Analyzers
Preliminary Checks
E8362A, E8363A, E8364A
• A repeated failure in
any
of the ranges of the attenuator check indicates a faulty step
attenuator.
— If the Port 1 check fails, the A25 step attenuator may be faulty.
— If the Port 2 check fails, the A26 step attenuator may be faulty.
Refer to
“Removing and Replacing the A36 and A37 60-dB Step Attenuators (Option
.
• If
all
ranges of the attenuator check fail, suspect a loose connection on the step
attenuator control cable.
— If the Port 1 check fails, check the A25 step attenuator cable.
— If the Port 2 check fails, check the A26 step attenuator cable.